Button Box Cleaning Instructions
Over time, residue may build up between the buttons and the button housings on the system. This can cause the buttons to stick slightly in the housing when pressed down and can slow the return of the buttons to the up position when released. Having the middle button in this condition can result in Task 3 in the assessment occasionally having the on-screen box advance through the moving vertical lines without client input. If this occurs, please clean the button according to the following instructions.
Problem:
During Task 3, the box will move even though the center button was only pressed once.
Solution:
The buttons needs to be cleaned with 99% Isopropyl alcohol.
Detailed Steps:
- You will need 99% isopropyl alcohol and cotton swabs.
- Power down and unplug the computer.
- Dip one end of the cotton swab in the alcohol. Rub the soaked end along the all side edges of the button.
- Depress the button into the button housing for a few seconds, and release.
- Take the dipped end of the cotton swab, rub the edge the buttons. Flip to the dry side of the swab, and repeat.
- Wait a few seconds to let the alcohol fully evaporate, and repeat the above steps a few more times, using a fresh swab each time.
- Power up the system, login to the DriveABLE Assessment site, and test the button using Task 3 in a Practice Assessment, found in the Training section of the site.
Should you have any questions or require additional assistance, you can contact us via one of the following methods:
- Submitting a ticket by using the "Add Ticket" button at the top of the page.
- Using the chat feature at the bottom right corner of this page.
- Sending us an email at [email protected]
- Giving us a call at 1-855-365-3748 ext. 1.
Related Articles
Installing a new Button Box
Problem: The button box is no longer functional or the cleaning instructions are no longer effective, you will have to contact Impirica support for a new button box. Once you receive the new button box, you can follow the following instructions to ...
Device Disinfection Protocol
To minimize the risks associated with secondary transmission of illness or infection, it is important to use disinfectant to clean the machine after each client. If the DCAT unit is not turned off, please use the Windows Shutdown option to power the ...
DCAT FAQ's
How long does the assessment take? The DriveABLE Cognitive Assessment takes approximately 30-60 minutes to complete. Some jurisdictions may require additional testing. Who facilitates the assessment? A trained and DriveABLE certified professional ...
Changing Computer/Monitor on the Stand
Problem: Your touch screen computer needs to be replaced and you have received a new device from Impirica. You can follow the below instructions to complete the replacement. Solution: Ensure the computer is powered down and disconnected from the wall ...
Connecting to a Hotspot
Setting up Personal Hotspot using an iPhone or iPad Settings Cellular Personal Hotspot or Settings Personal Hotspot Allow Others to Join Setting up Personal Hotspot using an Android device Settings Network and Internet Hotspot and ...